2025-06-24 @ ALS 8.3.1

An impromptu trip to ALS during Astera meeting. We collected diffraction data from several SARS-CoV-2 NSP3 macrodomain crystals at room temperature.

  • James, Mehagan, and Steve collected data from 4 macrodomain crystals harvested from a single tray "Plate 1" provided by JF's lab.
  • There was some messing around before we got the dose right. So each dataset had different beam properties (divergence & flux).
  • The crystals were of moderate size (~ 100 x 200 x 200 µm^3)
  • We used the usual Ando lab method: MiTeGen plastic sleeves, background measurements.
  • We did NOT use a humidity controlled environment for harvesting (not enough time to set it up).

Usable datasets:

file prefix tray, well dose rate* notes
H6_5 tray 1, well H6 high 360 degrees, 0.1 deg/frame, spots decayed rapidly (damage)
H8_8 tray 1, well H8 medium 360 degrees, 0.1 deg/frame, some damage toward the end?
G8_1 tray 1, well G8 low 360 degrees, 0.1 deg/frame

*by adjusting attenuators and mirror defocus, need to ask James H how to retrieve this info from metadata file